Contact Info
TEL: 028-84193018
FAX: 028-84193058
Add: 515,Building 1,NO39,Jiancai road,Chenghua district,Chengdu,Sichuan,China
Biochemistry Analyzer

Full Automatic Benchtop Biochemistry Analyzer YSD320

Item No.: YSD374
$0.00
INQUIRY
Description
YSD-320 Fully-auto Biochemistry Analyzer


Features
  • Random access, discrete automatic biochemistry analyzer, Insert emergency at any time
  • High performance and reasonable price, It is the good economic choice from semi-auto upgraded to fully-auto analyzer
  • Probe performance: Liquid level detection, all-round anti-collision protection and inventory checking, have automatic stop and alarm prompt for an obstacle
  • Probe cleaning: Automatic washing for both interior and exterior
  • Separate mixer, Auto washing, 24hours refrigeration system

Analysis System
  • Constant 200tests/hour, up to 320tests/hour
  • Test method: End-point, Fixed-time (two point method), kinetic
  • Reaction tray:  81 UV cuvettes, reusable
  • Mini reaction volume: 150μl, save reagent cost
  • Insert emergency at any time
  • Support linear and non-linear calibration
  • Heating method: solid constant temperature closed direct heating
  • Temperature equilibrium time: less than 3minutes,
  • Reaction Temperature: 37±0.2°C
  • Temperature Fluctuation: ± 0.1°C
 
Optical System

 
  • Integrated full closed optical system, no mechanical parts, not moldy, no maintenance cost
  • 8 wavelengths, selectable among 340-800nm, after spectrometric method
  • Absorption range: 0.00-4.00Abs
  • Halogen-tungsten lamp has more than 2000hrs service time
  • Center wavelength deviation: ±2nm, half wave width less than 12nm
  • Stray light less than 0.3%
  • Resolution: 0.0001Abs
 
Sample adding and rinsing system

 
  • Sample volume: 2- 100μl, 0.1μl by step
  • Reagent volume: 20-500μl, 1μl by step
  • Water consumption: less than 5L/hour
 
 
Mixing and washing system
  • Mixing system:               separate mixer, to mix the sample and reagent completely, for better reaction effect
  • Washing system:     single needle to wash cuvettes one by one automatically
 
Reagent/Sample Tray

  • 60 reagent + 60 sample positions
  • Advanced reagent tray refrigeration system, temperature: 4-12°C
  • Automatically detect reagent remain, real time display reagent remain
  • Reagent bottles with 15ml and 30mlare available, dead volume<0.5ml
  • Support single/ dual reagents, reagents are open
  • Real time display reagent tray status
  • Automatically judge substrate depletion and automatically dilute and retest
 
Operation System


  • Windows XP or Windows 7 Home, easy to operate
  • Set the operating authority, no time limit and storage limit,
  • Permanently preserve all kinds of data
  • Automatic storage, automatic backup,
  • English print repo rt, the user can modify the print format
  • English online help and operation guide
  • The test parameter setting is completely open
  • The unique user password management, reasonable distribute the operating authority
  • Real-time display the status of sample testing, reagent remain, reaction tray and reaction curve recording
  • Real-time understand the testing progress
  • Automatically check calibration curve and fitting the best calibration curve
  • Calibration Mode: Linear, Logit-4p, Logit-5p, Spline, Factor, Polynomial
  • The computing items can be set randomly, automatically judge substrate depletion, automatically retest
  • Real-time display the reaction tray temperature
  • RS232 standard interface, support the intranet and remote printing

Technical Specifications      
                                                 
Performance

Type: Discrete Automatic, Random Access
Principle:  Absorption Spectrometry, Turbidimetry
Throughput:  Constant 200tests/hour (without ISE)
Test Method:  End-point, Fixed-time, kinetic
Test Sequence:  Run tests in strict accordance with sample sequence,
                         Insert emergency at any time
 

Optical System

Light Source: Halogen-tungsten lamp, 12V/20W More than 2000hrs service time
Spectrometric Method:  After spectrometric method
Wavelength: 8 wavelengths, selectable among 340-800nm
Detector: Photodiode array
Absorption Range: 0-4.0Abs
Optical System: Integrated full closed optical system
 

Reaction system

Reaction Tray:  81 positions
Reaction Volume:  Mini 150μl
 

Sample system

Sample Tray: 60 positions
Sample Volume: 2-100μl , 0.1μl by step
Online Sample: 60 samples/ batch, unlimited batch input
 

Reagent System

Reaction Tray: 60 positions, open system
Reagent Volume: 20-500μl, 1μl by step
Reagent Bottle: Bottles with 15ml and 30ml are available, dead volume<0.5ml
 

Temperature Control System

Heating Mode: Solid constant temperature closed direct heating
Heating Time: less than 3minutes
Reaction Temperature:  37±0.2°C
Temperature Fluctuation: ± 0.1°C
Refrigeration system: 24 hours cooling for reagent tray, temperature: 4-12°C
 

Calibration and QC

Calibration Mode:  Linear, Logit-4p, Logit-5p, Spline, Factor, Polynomial
                            Automatically check calibration curve
and fitting the best calibration curve
QC Type: Multi-type
QC Rules: Multi-rule; default: Westguard
Status Monitoring:  Real -time monitor the sample tray, reagent tray, reaction tray; Real-time display reaction tray humidity, reagent remain, reaction curve, calibration curve and QC chart.
Inspection and Judgment: Reaction end point judgment, reaction linear range judgment, depletion judgment, reagent blank absorbance judgment

Operation and Print

Operation System: Windows XP or Windows 7 Home
Database: SQL Server
Data Storage: Automatic storage and backup, permanently preserve all kinds of data
Print Content: Print content can be customized
Print Format: Provide a variety of print format
 

Working Conditions

Water Consumption:  less than 5L/hour
Power Supply: AC 220 ±22V  50HZ ±1HZ
Power: Max 200VA
Ambient Temperature: 15°C-30°C
Dimension(L*W*H): 79*59*108cm
Net Weight: 90KGS
 
TOP