Model No. | YSD186B |
Infusion Mode | Volumetric |
Pumping Mechanism | Peristaltic finger pump |
IV Sets | Standard IV sets of 10, 15, 20 and 60d/mL, with diameter 3.4 ~ 4.5mm. |
FlowRate Range | 1.0 ~ 999.9 mL/h |
Increment | 0.1 mL/h |
Flow Rate Accuracy | ≤ ± 5% (after calibration) |
Volume Limit | 1~9999 mL |
Volume Infused Display | 0~9999 mL |
Purge Rate | 600 mL/h |
KVO Rate | 1 mL/h |
Occlusion Pressure | Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa. |
Air-in-line Detection | Air bubble detected by ultrasonic and the device with 8 sensitivities setting,50 ~ 1000μL. |
Alarm Volume | 8 levels setting |
IV Set Management and Flow Rate Calibration | Up to 10 adjustment values for various brands IV set could be stored to make the operation more convenient. |
Anti-reve rse Function | Anti-reve rse function on motor to prevent upstream |
Self-check | Self-check when start up and during infusion |
Audio and Visual Alarms | Air-in-line, Door open, Occlusion, Infusion finished, No operation, Battery low, Pump error, Motor reve rse, Pressure abnormity, Sound abnormity, AC power lost. |
Display | Color LCD with high brightness |
Power Source | AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA |
Battery | Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h |
Operating Conditions | Temperature: 5°C ~ 40°C Relative humidity: 20% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Storage Conditions | Temperature: -20°C ~ 55°C Relative humidity: 10% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Classification | Class II, CF, IPX4 |
Dimensions and N.W. | 150mmx 150mm x 235mm; 2.3kg |
Application | Intravenous infusion for various liquid medicines, such as vascular expansion solution, anticancer drug, accelerative procreates drug, liquid nutrient, etc. Applicable for micro and rapid infusion. |